BQG-2118 Industrial Online Zirconia Oxygen Analyzer

Technical indexes


1.Measuring range: 0~20.6%02

2.Accuracy: ≤±2%FS

3.Temperature of flue gas measured: ≤600℃(low temperature type)600~800(high temperature type)

4. Excess air coefficient of flue gas measured: 0.1-99                    

5. Response time: ≤ 3 seconds (90% response)

6. Current output: 0~10mADC or 4-20mADC

7. Power supply: 220V±10% 50HZ± 5%

8.Working conditions: ambient temperature: 0~50  relative humidity:<85%

9.Heating-up time of the heater of zirconia detector: Approx 20 minutes  

10.Overall dimension and dimension of the hole



Type

Overall dimension mm

Dimension of the hole mm (height * width)

Horizontal (H)

160x80x120

152+1×76+1

Vertical (Z)

80x160x120

   76+1×152+1

Square (Q)

160x160x200

    152+1×152+1

Wall mounting (G)

220 x 170 x 75 (height x width x thickness)

BQG-2118 zirconia oxygen analyzer is practical and reliable on-line automatic oxygen analyzer. Its measurement and display system consists of an oxygen detector (sensor) and a display instrument (transmitter). The detector is made of heat-resisting and anti-corrosion materials and can be directly inserted into flues for measurement of oxygen. The display instrument adopts single-chip control and can be used in combination with various electric unit combination instrument, main display recorder instruments and distributed control systems (DCS).

Zirconia oxygen analyzer is used for rapid and accurate on-line measurement and analysis of the oxygen contents in the flue gas produced by combustion equipment, such as boilers, furnaces and heaters, so as to realize control of low-oxygen combustion, energy saving and pollution treatment. 

Features


The sensor is made of heat-resisting and anti-corrosion materials with high reliability.

With a special structural design, this analyzer doesnt need external air pump, and the reference air can realize 

convection by itself.

The sensor has a standard air interface, which enables calibration check on the site and provides convenient

maintenance.

It adopts single-chip control and data processing with quick response, accurate measurement and stable

performance.

上海博取儀器有限公司
  • 聯(lián)系人:邢銀鳳
  • 地址:上海浦東新區(qū)秀沿路118號
  • 郵箱:2355456200@qq.com
  • 手機:15021885022
關(guān)注我們

歡迎您關(guān)注我們的微信公眾號了解更多信息

掃一掃
關(guān)注我們
版權(quán)所有©2025上海博取儀器有限公司All Rights Reserved    備案號:滬ICP備08017338號-29    sitemap.xml    總訪問量:672041
管理登陸    技術(shù)支持:儀表網(wǎng)
北票市| 佛坪县| 广元市| 亳州市| 岳池县| 呈贡县| 吴堡县| 长垣县| 诸暨市| 浮梁县| 高雄县| 禄丰县| 南昌市| 石阡县| 连江县| 永靖县| 杨浦区| 甘泉县| 衢州市| 南澳县| 当雄县| 疏勒县| 界首市| 原平市| 保亭| 南开区| 论坛| 蛟河市| 克山县| 兴国县| 梓潼县| 长春市| 德令哈市| 都兰县| 乐业县| 江源县| 夏邑县| 台州市| 天全县| 二连浩特市| 荃湾区|